Sterilization Methods for Medical Plastic Bottles: Gamma Radiation vs. Ethylene Oxide

Sterilization Methods for Medical Plastic Bottles

Medical plastic bottles require sterilization to meet aseptic processing requirements. This article compares sterilization methods and their impact on PET materials.

Sterilization Methods Comparison

MethodDose/ConditionsPET Compatibility
Gamma Radiation25-40 kGyGood (minor yellowing)
E-Beam25-30 kGyExcellent
EtO (Ethylene Oxide)450-600 mg/L, 37-55°CExcellent
Steam (Autoclave)121°C, 15 psi, 30 minPoor (deformation)

Gamma Radiation Effects

  • Dose Range: 25 kGy (sterility assurance level 10⁻⁶)
  • Color Change: Δb-value +2 to +5 (slight yellowing)
  • Mechanical Properties: ≤10% change in tensile strength
  • Molecular Weight: Minor chain scission, IV reduction ≤0.05 dL/g

EtO Sterilization

  • Cycle time: 12-18 hours including aeration
  • Residual EtO: ≤1 ppm (ISO 10993-7)
  • No material degradation
  • Excellent for heat-sensitive products

Aseptic Processing Requirements

✓ ISO 13485 medical device quality management
✓ Clean room filling (ISO Class 7 or better)
✓ Sterility testing per USP <71>
✓ Endotoxin testing: ≤0.5 EU/mL

Validation Documentation

Sterilization validation per ISO 11137 (radiation) or ISO 11135 (EtO) with dose mapping and biological indicators.


📱 WhatsApp: +86 181 5928 4005
✉️ Contact us for technical consultation and product specifications!

See our pharmaceutical bottles and HDPE bottles, or request a quote for medical-grade packaging.

Obtenez un devis

Tell us capacity, material and quantity
We come back with options and pricing

Contact WhatsApp
Code QR WeChat Identifiant WeChat +8618159284005
E-mail +86 181 5928 4005